Comparative analysis of lateral cervical radiographic imaging quality with and without anti-scatter grid application
ID Lučić, Daša (Avtor), ID Jurša, Laura (Avtor), ID Alukić, Erna (Avtor), ID Mekiš, Nejc (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (489,02 KB)
MD5: DDE4433F73FBAE3F69C4FBCEB15220C6
URLURL - Izvorni URL, za dostop obiščite https://www.sciencedirect.com/science/article/pii/S0969804326000436 Povezava se odpre v novem oknu

Izvleček
Purpose This study investigated the effects of anti-scatter grid removal on radiation dose, image quality, and anatomical visibility in lateral (LAT) cervical spine imaging. Methods A two-part prospective study was conducted. In the phantom study, an anthropomorphic head-torso phantom with lard simulating extra fatty tissue of 2–10 cm in 2 cm increments was imaged using protocols with and without an anti-scatter grid. In the clinical study, 80 patients were randomly divided into two groups: half were imaged with an anti-scatter grid, and half without. Data collected included gender, body mass index (BMI), Dose Area Product (DAP), tube voltage (kV), current-time product (mAs), filtration, neck width, and use of Automatic Exposure Control (AEC). Effective dose (ED) and dose to selected organs were calculated using PCXMC 2.0 software. Two traumatologists and one radiologist evaluated all phantom and patient images according to European guidelines using a four-point scale (1 = diagnostically inadequate image; 4 = excellent image quality). Results In the phantom study, removal of the anti-scatter grid reduced mean DAP by 52.1 % (p < 0.001) without affecting diagnostic adequacy; soft tissue-bone contrast improved (p = 0.009). In the patient study, removal of the anti-scatter grid reduced mean DAP and ED by 63.1 % and 61.7 %, respectively (both p < 0.001). All organ doses were significantly lower (p < 0.001). Soft tissue visibility improved by 12.5 % (p = 0.011), and soft tissue-bone contrast improved by 9.1 % (p = 0.007) without the anti-scatter grid. No other quality criteria differed significantly. Conclusion In both phantom and patient studies, removal of the anti-scatter grid resulted in significant dose reduction while maintaining or improving image quality. These results are consistent with previous findings and support the As Low As Reasonably Achievable (ALARA) principle. Omitting the anti-scatter grid in LAT cervical spine imaging reduces dose by about 60 % without compromising diagnostic quality. Routine use in clinical practice is recommended to increase patient safety.

Jezik:Angleški jezik
Ključne besede:cervical spine radiography, dose reduction, image quality, anti-scatter grid
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:ZF - Zdravstvena fakulteta
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Datum objave:01.04.2026
Leto izida:2026
Št. strani:5 str.
Številčenje:Vol. 230, art. 112459
PID:20.500.12556/RUL-178930 Povezava se odpre v novem oknu
UDK:616-07
ISSN pri članku:1872-9800
DOI:10.1016/j.apradiso.2026.112459 Povezava se odpre v novem oknu
COBISS.SI-ID:266902019 Povezava se odpre v novem oknu
Datum objave v RUL:02.02.2026
Število ogledov:355
Število prenosov:264
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:Applied radiation and isotopes
Založnik:Elsevier
ISSN:1872-9800
COBISS.SI-ID:23195141 Povezava se odpre v novem oknu

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.

Podobna dela

Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:

Nazaj