Podrobno

Development and validation of an argon triple point apparatus with a novel automatic pressure control system
ID Matas, Ivan (Avtor), ID Grgec Bermanec, Lovorka (Avtor), ID Šestan, Danijel (Avtor), ID Bojkovski, Jovan (Avtor), ID Žužek, Vincencij (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (13,90 MB)
MD5: BD27227CCA6E80DC749B3F1F5D24C599
URLURL - Izvorni URL, za dostop obiščite https://www.mdpi.com/1424-8220/25/5/1411 Povezava se odpre v novem oknu

Izvleček
This paper describes the development and validation of an apparatus for the realization of the triple point of argon (83.8058 K), with a novel automatic pressure control system for the liquid nitrogen cryostat. The automatic pressure controller, together with custom-made software, was developed and tested in the Laboratory for Process Measurement at the Faculty of Mechanical Engineering and Naval Architecture, University of Zagreb (FSB-LPM). Performance testing and characterization of the automatic pressure controller confirmed its suitability for precise and reliable control of gauge pressure in the cryostat. The characteristics and uncertainty of the measurement setup for the realization of the triple point of argon were validated through a bilateral hybrid comparison with the Laboratory of Metrology and Quality at the Faculty of Electrical Engineering, University of Ljubljana (MIRS/UL-FE/LMK). A long-stem quartz-sheathed standard platinum resistance thermometer was used as a transfer standard. The realizations of the International Temperature Scale (ITS-90) were compared in the subrange from the triple point of argon to the triple point of water. The comparison results show that resistance ratio (W) values determined by FSB-LPM at the fixed points of argon and mercury deviate from the MIRS/UL-FE/LMK values, within the determined combined uncertainty of the comparison.

Jezik:Angleški jezik
Ključne besede:triple point of argon, automatic pressure control, bilateral comparison, standard platinum resistance thermometer, International Temperature Scale of 1990
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:FE - Fakulteta za elektrotehniko
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Leto izida:2025
Št. strani:22 str.
Številčenje:Vol. 25, iss. 5, art. 1411
PID:20.500.12556/RUL-167798 Povezava se odpre v novem oknu
UDK:621.317:006.91
ISSN pri članku:1424-8220
DOI:10.3390/s25051411 Povezava se odpre v novem oknu
COBISS.SI-ID:228749571 Povezava se odpre v novem oknu
Datum objave v RUL:12.03.2025
Število ogledov:78
Število prenosov:11
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:Sensors
Skrajšan naslov:Sensors
Založnik:MDPI
ISSN:1424-8220
COBISS.SI-ID:10176278 Povezava se odpre v novem oknu

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:trojna točka argona, avtomatska regulacija tlaka, medlaboratorijska primerjava, standardni platinasti uporovni termometer, Mednarodna temperaturna lestvica iz 1990