Veljavnost in ponovljivost EasyForce dinamometra na primeru kolena in kolka : magistrsko delo
ID Križaj, Luka (Avtor), ID Šarabon, Nejc (Mentor) Več o mentorju... Povezava se odpre v novem oknu

.pdfPDF - Predstavitvena datoteka, prenos (1,61 MB)
MD5: 164998C85E8EBC5951AAD2B8C8CCBB94

Izvleček
Namen naloge je bil preveriti veljavnost ter ponovljivost rezultatov (znotraj obiska in med obiski), pridobljenih z ročnim dinamometrom EasyForce, upoštevajoč postopke, ki so v skladu s priporočili proizvajalca, na primeru kolena in kolka. V raziskavo je bilo vključenih 50 mladih, zmerno treniranih odraslih. Veljavnost in ponovljivost sta bili testirani s parnim dvo-repim t-testom, katerega velikost učinka smo izrazili kot Cohenov d (d), s Pearsonovim koeficientom korelacije (r), intra-klasnim koeficientom korelacije (ICC) in koeficientom variacije (KV). Ugotovili smo, da med vrednostmi, izmerjenimi z EasyForce dinamometrom in referenčno metodo z namenskima dinamometroma z zunanjo pritrditvijo, prihaja do statistično značilnih razlik (p < 0,001). Vrednosti iztegovalk in upogibalk kolena, izmerjene z Easyforce dinamometrom, so približno dvakrat nižje od vrednosti, ki so bile izmerjene z referenčno metodo z namenskim dinamometrom z zunanjo fiksacijo (-1,88 ≤ d ≤ -2,02), medtem ko so vrednosti odmikalk in primikalk kolka približno dva- do tri-krat višje (1,81 ≤ d ≤ 2,91). Korelacija med vrednostmi je v celoti močna pri primikalkah kolka (r ≥ 0,70) in zmerna do močna pri iztegovalkah in upogibalkah kolena ter odmikalkah kolka (0,64 ≤ r ≤ 0,77). Relativna ponovljivost meritev jakosti z EasyForce-om znotraj obiska je bila dobra pri iztegu kolena (0,84 ≤ ICC ≤ 0,88) in odlična pri vseh ostalih mišičnih skupinah (ICC ≥ 0,90). Absolutna ponovljivost znotraj obiska je bila sprejemljiva pri meritvah največje jakosti in največje hitre jakosti upogibalk kolena in odmikalk kolka (KV ≤ 9,90%) ter pri meritvah primikalk kolka na drugem obisku (KV ≤ 9,60%). Pri vseh meritvah iztegovalk kolena in primikalk kolka na prvem obisku je bila absolutna ponovljivost nesprejemljiva (KV > 10,0%). Relativna ponovljivost meritev z EasyForce-om med obiski je bila, ne glede na nalogo, odlična (ICC ≥ 0,95), absolutna ponovljivost pa sprejemljiva (KV ≤ 7,70%).

Jezik:Slovenski jezik
Ključne besede:ročni dinamometer, veljavnost, ponovljivost, spodnje okončine, mišična jakost
Vrsta gradiva:Magistrsko delo/naloga
Tipologija:2.09 - Magistrsko delo
Organizacija:FŠ - Fakulteta za šport
Leto izida:2021
PID:20.500.12556/RUL-129557 Povezava se odpre v novem oknu
COBISS.SI-ID:81484035 Povezava se odpre v novem oknu
Datum objave v RUL:05.09.2021
Število ogledov:812
Število prenosov:166
Metapodatki:XML RDF-CHPDL DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Sekundarni jezik

Jezik:Angleški jezik
Naslov:Validity and repeatability of EasyForce dynamometer in the case of knee and hip
Izvleček:
The purpose of the study was to test validity and (intra- and inter-sessions) repeatability of the results obtained by the hand-held dynamometer EasyForce, following procedures that are in accordance with manufacturer’s recommendations, in the case of the knee and hip. The study included 50 moderately trained young adults. Validity and repeatability were statistically tested with a paired two-tailed t-test, whose magnitude of effect was expressed as Cohen’s d (d), with Pearson’s correlation coefficient (r), intra-class correlation coefficient (ICC) and coefficient of variation (CV). We found out there are statistically significant differences between results measured with EasyForce hand-held dynamometer and reference method with dedicated dynamometer with external attachment (p < 0,001). The values of knee extensors and flexors measured with EasyForce dynamometer were approximately twice lower than the values measured with reference method with dedicated dynamometer with external attachment (-1,88 ≤ d ≤ -2,02), while the values of hip abductors and adductors were approximately two to three times higher (1,81 ≤ d ≤ 2,91). Correlation between the values was completely strong for hip adductors (r ≥ 0,70) and moderate to strong for knee extensors, knee flexors and hip abductors (0,64 ≤ r ≤ 0,77). Relative intra-session repeatability of strength measurements with EasyForce was good for knee extension (0,84 ≤ ICC ≤ 0,88) and excellent for all other muscle groups (0,90 ≤ ICC ≤ 0,96). Absolute intra-session repeatability was acceptable for measurements of maximal muscle strength and rate of force development of knee flexors and hip abductors (CV ≤ 9,90%) and for measurements of hip adductors on the second visit (CV ≤ 9,60%). Absolute intra-session repeatability was unacceptable for all measurements of knee extensors and for measurements of hip adductors on first visit (CV > 10,0%). Relative inter-sessions repeatability of EasyForce measurements was, regardless of the task, excellent (ICC ≥ 0,95) and absolute inter-sessions repeatability was acceptable (CV ≤ 7,70%).

Ključne besede:hand-held dynamometer, validity, repeatability, lower extremity, muscle strength