Comparison of tandem mass spectrometry and the fluorometric method : parallel phenylalanine measurement on a large fresh sample series and implications for newborn screening for phenylketonuria
ID Perko, Daša (Avtor), ID Grošelj, Urh (Avtor), ID Čuk, Vanja (Avtor), ID Remec, Žiga Iztok (Avtor), ID Žerjav-Tanšek, Mojca (Avtor), ID Drole Torkar, Ana (Avtor), ID Krhin, Blaž (Avtor), ID Biček, Ajda (Avtor), ID Oblak, Adrijana (Avtor), ID Battelino, Tadej (Avtor), ID Repič-Lampret, Barbka (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (2,70 MB)
MD5: C1E659ED8BF9FFCCB6ED8F1E0EE6504E
URLURL - Izvorni URL, za dostop obiščite https://www.mdpi.com/1422-0067/24/3/2487 Povezava se odpre v novem oknu

Izvleček
Phenylketonuria (PKU) was the first disease to be identified by the newborn screening (NBS) program. Currently, there are various methods for determining phenylalanine (Phe) values, with tandem mass spectrometry (MS/MS) being the most widely used method worldwide. We aimed to compare the MS/MS method with the fluorometric method (FM) for measuring Phe in the dried blood spot (DBS) and the efficacy of both methods in the NBS program. The FM was performed using a neonatal phenylalanine kit and a VICTOR2$^{TM}$ D fluorometer. The MS/MS method was performed using a NeoBase$^{TM}$ 2 kit and a Waters Xevo TQD mass spectrometer. The Phe values measured with the MS/MS method were compared to those determined by the FM. The cut-off value for the NBS program was set at 120 µmol/L for FM and 85 µmol/L for MS/MS. We analyzed 54,934 DBS. The measured Phe values varied from 12 to 664 µmol/L, with a median of 46 µmol/L for the MS/MS method and from 10 to 710 µmol/L, with a median of 70 µmol/L for the FM. The Bland–Altman analysis indicated a bias of −38.9% (−23.61 µmol/L) with an SD of 21.3% (13.89 µmol/L) when comparing the MS/MS method to the FM. The Phe value exceeded the cut-off in 187 samples measured with FM and 112 samples measured with MS/MS. The FM had 181 false positives, while the MS/MS method had 106 false positives. Our study showed that the MS/MS method gives lower results compared to the FM. Despite that, none of the true positives would be missed, and the number of false-positive results would be significantly lower compared to the FM.

Jezik:Angleški jezik
Ključne besede:tandem mass spectrometry, MS/MS, fluorometric method, FM, comparison, phenylalanine, Phe, phenylketonuria, PKU, newborn screening, NBS, false positive results, recall rate
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:MF - Medicinska fakulteta
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Leto izida:2023
Št. strani:11 str.
Številčenje:Vol. 24, iss. 3, art. 2487
PID:20.500.12556/RUL-155551 Povezava se odpre v novem oknu
UDK:616-053.2
ISSN pri članku:1422-0067
DOI:10.3390/ijms24032487 Povezava se odpre v novem oknu
COBISS.SI-ID:147446787 Povezava se odpre v novem oknu
Datum objave v RUL:05.04.2024
Število ogledov:152
Število prenosov:12
Metapodatki:XML RDF-CHPDL DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:International journal of molecular sciences
Skrajšan naslov:Int. j. mol. sci.
Založnik:MDPI
ISSN:1422-0067
COBISS.SI-ID:2779162 Povezava se odpre v novem oknu

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:tandemska masna spektrometrija, MS/MS, fluorometrična metoda, FM, primerjava, fenilalanin, Phe, fenilketonurija, pregled novorojenčka, PKU, lažno pozitivni rezultati, NBS, stopnja odpoklica

Projekti

Financer:ARRS - Agencija za raziskovalno dejavnost Republike Slovenije
Številka projekta:P3-0343
Naslov:Etiologija, zgodnje odkrivanje in zdravljenje bolezni pri otrocih in mladostnikih

Podobna dela

Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:

Nazaj