izpis_h1_title_alt

Haptični tetivni sistem za urjenje ravnotežja med hojo
ID KOVŠE, JAKA (Avtor), ID Mihelj, Matjaž (Mentor) Več o mentorju... Povezava se odpre v novem oknu, ID Olenšek, Andrej (Komentor)

.pdfPDF - Predstavitvena datoteka, prenos (40,06 MB)
MD5: E63772F160D3B5CF5CE54ED03E70F266

Izvleček
Hoja je kompleksen način premikanja, saj morajo za učinkovito izvajanje pravilno delovati živčni sistem, mišično-skeletni sistem in kardiovaskularni sistem. Okvare teh sistemov lahko povzročajo ravnotežne motnje in motnje pri hoji, ki posledično vodijo k poškodbam povezanimi s padci ali celo smrti. Glavni povzročitelji motenj so staranje, bolezni srca in ožilja, ortopedski defekti in nevrološke bolezni. Zaradi staranja populacije in večjih možnosti za preživetje nevroloških in kardiovaskularnih bolezni se potreba po rehabilitaciji hoje veča. Posledično se stroški izvajanja manualne terapije hoje in ravnotežja večajo, obenem pa primanjkuje kadra. Zato prihaja do razvoja rehabilitacijskih naprav, ki fizioterapevtom omogočajo lažji, bolj ponovljiv in učinkovitejši način treninga hoje. Večina rehabilitacijskih naprav na trgu naslavlja problematiko patološke hoje iz vidika izboljšanja mehanizmov hoje, ne obravnava pa problematike ravnotežja med hojo in ponovne vzpostavitve refleksnih odzivov za preprečitev padcev. V ta namen se razvija nov pristop rehabilitacije hoje, ki temelji na izvajanju nadzorovanih perturbacij v varnem in nadzorovanem okolju. Na Univerzitetnem inštitutu za rehabilitacijo Republike Slovenije - SOČA so v ta namen razvili paralelni robotski mehanizem za ocenjevanje ravnotežja (BAR). V okviru magistrske naloge smo obravnavali problematiko zmanjšanja negativnega vpliva vztrajnosti paralelnega mehanizma na dinamiko uporabnika, ki je glavna slabost sistema BAR. Razvili smo tetivni sistem, ki bi na medenico osebe vsiljeval nadzorovane perturbacije v eni prostostni stopnji, lahko pa bi nudil pomoč med gibanjem in spodbujal učenje ravnotežja. Razvit sistem je sestavljen iz po meri narejene konstrukcije, ki povezuje dva pogonska dela, vsakega na eni strani platforme, na kateri stoji uporabnik. Posamezen pogonski del vleče vrvico, ki poteka po sistemu škripčevja od pogonskega dela do pasu, s katerim je v sistem vpet uporabnik. Vodenje motorjev in izračun prikaza navideznega okolja poteka na programirljivem logičnem krmilniku. Po preučitvi zahtev smo se odločili za uporabo zaprtozančnega admitančnega vodenja. Delovanje tetivnega sistema smo razdelili na proces inicializacije, s katerim smo izpolnili zahtevane začetne pogoje tetivnega sistema in na prosto delovanje, v okviru katerega smo generirali različne profile zunanjih sil na uporabnika, omogočali prosto gibanje in generirali polje sil. Delovanje tetivnega sistema smo ovrednotili z rezultati transparentnosti sistema, z rezultati zmožnosti generiranja polja sil v eni prostostni stopnji in z rezultati zmožnosti generiranja različnih profilov perturbacij. Zadovoljiv prikaz praznega prostora smo dosegli z vpeljavo kompenzacije lastne dinamike tetivnega sistema in preučili njen vpliv na zmožnost generiranja interakcijskih sil navideznega okolja. Ugotovili smo, da naš način kompenzacije poslabša zmožnost generiranja sil interakcije navideznega okolja, zato lahko v vodenje vpeljemo dodatno logiko spreminjanja vpliva kompenzacije. Tetivni sistem je zagotavljal generiranje ustreznega polja sil, ki uporabniku onemogoča pretiran odmik od želene lege in na ta način nudi pomoč pri hoji. Zaradi omejitev napajalnega sistema pa je tetivni sistem omogočal generiranje 100 ms kratkih impulzov sil le do amplitude 85 N. V okviru tega dela smo izdelali tetivni sistem za vodenje človekove medenice v eni prostostni stopnji. Z rezultati meritev ter dejstvom, da bi lahko zmogljivost sistema izboljšali z uporabo namenskih in kakovostnejših komponent smo podkrepili smiselnost nadaljnjega razvoja. Uporaba škripčevja omogoča veliko število različnih konfiguracij, sistem pa obenem lahko razširimo z dodatnimi aktuatorji. Strojna širitev sistema je torej trivialna, poleg tega pa uporaba tetivnih sistemov omogoča zmanjšanje vpliva lastne dinamike sistema na gibanje uporabnika. Tega za izdelavo podobnega sistema z uporabo togega paralelnega mehanizma ne moremo trditi.

Jezik:Slovenski jezik
Ključne besede:rehabilitacija, tetivni sistem, perturbacija, admitančno vodenje
Vrsta gradiva:Magistrsko delo/naloga
Organizacija:FE - Fakulteta za elektrotehniko
Leto izida:2022
PID:20.500.12556/RUL-137613 Povezava se odpre v novem oknu
COBISS.SI-ID:113028611 Povezava se odpre v novem oknu
Datum objave v RUL:23.06.2022
Število ogledov:482
Število prenosov:90
Metapodatki:XML RDF-CHPDL DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Sekundarni jezik

Jezik:Angleški jezik
Naslov:Haptic tendon-driven system for balance training during walking
Izvleček:
Walking is a complex way of moving, as the nervous system, musculoskeletal system and cardiovascular system must function properly to perform effectively. Any failure of these systems can cause balance and gait disturbances, which in turn lead to injuries related to falls or even death. The main causes of balance and gait disturbances are aging, cardiovascular disease, orthopedic defects and neurological disease. Due to the aging population and increased chances of survival of neurological and cardiovascular diseases, the need for rehabilitation is increasing. As a result, the costs of performing manual walking training and balance therapy are increasing, and at the same time a shortage of staff is happening. Therefore, there is a tendency to develop rehabilitation devices that allow physiotherapists to make walking easier, more repeatable and more effective. Most rehabilitation devices on the market address the issue of pathological walking in terms of improving the mechanisms of walking, but do not address the issue of reactive balance control to prevent falls. To this end, a new approach to walking rehabilitation is being developed, based on the implementation of controlled perturbations in a safe and controlled environment. For this purpose, the University Institute for Rehabilitation of the Republic of Slovenia - SOČA has developed a parallel robotic balance assessment mechanism (BAR). In the master's thesis we discussed the problem of reducing the negative impact of the persistence of the parallel mechanism on the dynamics of the user, which is the main weakness of BAR. We developed a cable system that would impose controlled perturbations on users pelvis in one degree of freedom, but could help during movement and promote balance learning. The developed system consists of a custom-made construction that connects two drive parts, each on one side of the platform on which the user stands. An individual drive part pulls a cable that runs along the pulley system from the drive part to the belt with which the user is fastened to the system. Motor control and the calculation of the virtual environment interaction is running on a programmable logic controller. After examining the requirements, we decided to use closed-loop admittance control. The operation of the cable system was divided into the initialization process, which met the required initial conditions of the cable system and free operation, in which we generated various profiles of external forces on the user, allowed free pelvis movement and generated a force field. The operation of the tendon system was evaluated with the results of the transparency of the system, with the results of the ability to generate a force field in one degree of freedom and with the results of the ability to generate different perturbation profiles. A satisfactory display of empty space was achieved by introducing compensation of cable system dynamics and studying its influence on the ability to generate interactive forces of the virtual environment. We found that our method of compensation impairs the ability to generate the interaction forces of the virtual environment, so we can introduce additional logic into the management of changing the impact of compensation. The tendon system provided the generation of an appropriate force field, which prevents the user from excessive deviation from the desired position and thus provides assistance in walking. Due to the limitations of the power supply system, the string system allowed the generation of 100 ms short pulses of forces only up to the amplitude of 85 N. As part of this work, we developed a tendon system to guide the human pelvis in one degree of freedom. We supported further development with the results of measurements and the fact that the performance of the system could be improved by using dedicated and higher quality components. The use of pulleys allows a large number of different configurations, and the system can be extended with additional actuators. The mechanical expansion of the system is therefore trivial, and in addition, the use of cable systems makes it possible to reduce the influence of the system's own dynamics on the user's movement. This cannot be argued for making a similar system using a rigid parallel mechanism.

Ključne besede:rehabilitation, cable robot, perturbation, admittance control