Določanje homogenosti in stabilnosti tokovnega polja v merilni sekciji vetrovnika
ID Hrušovar, Nika (Avtor), ID Bobovnik, Gregor (Mentor) Več o mentorju... Povezava se odpre v novem oknu, ID Kutin, Jože (Komentor)

.pdfPDF - Predstavitvena datoteka, prenos (3,46 MB)
MD5: 922414731B6001D02812E40C50F1020F

Izvleček
Homogenost in stabilnost hitrostnega polja v merilni sekciji vetrovnika imata velik vpliv na rezultate merjenja hitrosti zraka v vetrovniku. V okviru magistrskega dela smo postavili in opisali novo merilno progo z dvo-komponentnim LDA merilnikom in vetrovnikom. Posebno pozornost smo posvetili popisu funkcij v nadzornem programu. Dobljeni rezultati kažejo, da je stabilnost pri hitrostih zraka med 10 in 30 m/s manj kot 0,2 %, pri hitrosti zraka 0,3 m/s pa je bila izračunana stabilnost več kot 10 %. Karakterizacija hitrostnih profilov je pokazala spreminjanje hitrostnega profila vzdolž merilne sekcije, kar je posledica geometrijske razširitve. Homogenost hitrostnega polja smo izračunali za različne velikosti pretočnih prerezov in različne hitrosti. Izračunana homogenost je manj kot 1,2 %, razen pri majhnih hitrostih (npr. 0,3 m/s), kjer znaša 3,8 %. Iz meritev homogenosti ne moremo izločiti vpliva nestabilnosti. Z izvedbo dodatne meritve smo opozorili tudi na upoštevanje zastojnega učinka pri umerjanju v polju pred merilnikom v merilni sekciji.

Jezik:Slovenski jezik
Ključne besede:Merjenje hitrosti, Laser Dopplerjev anemometer (LDA), hitrostni profil, vetrovnik, homogenost, stabilnost
Vrsta gradiva:Magistrsko delo/naloga
Tipologija:2.09 - Magistrsko delo
Organizacija:FS - Fakulteta za strojništvo
Kraj izida:Ljubljana
Založnik:[N. Hrušovar]
Leto izida:2019
Št. strani:XXII, [56] str.
PID:20.500.12556/RUL-109844 Povezava se odpre v novem oknu
UDK:533.6.071:551.508.54(043.2)
COBISS.SI-ID:16919067 Povezava se odpre v novem oknu
Datum objave v RUL:09.09.2019
Število ogledov:1638
Število prenosov:347
Metapodatki:XML DC-XML DC-RDF
:
HRUŠOVAR, Nika, 2019, Določanje homogenosti in stabilnosti tokovnega polja v merilni sekciji vetrovnika [na spletu]. Magistrsko delo. Ljubljana : N. Hrušovar. [Dostopano 30 marec 2025]. Pridobljeno s: https://repozitorij.uni-lj.si/IzpisGradiva.php?lang=slv&id=109844
Kopiraj citat
Objavi na:Bookmark and Share

Sekundarni jezik

Jezik:Angleški jezik
Naslov:Assessment of homogeneity and stability of the flow field in the measuring section of the wind tunnel
Izvleček:
The homogeneity and stability of the velocity field in the measuring section of the wind tunnel have a significant influence on the results of air velocity measurements in the wind tunnel. In this master's thesis, we set up and described a new measuring system with the two-component LDA and the wind tunnel. Special attention was devoted to the description of the functions in the control program. The results show that the stability of the air velocity between 10 and 30 m/s is less than 0.2 % and for lower velocities higher than 10 %. Characterization of velocity profiles showed a change in velocity profile along the measurement section resulting from the geometric extension. The homogeneity of the velocity field was calculated for different cross-sectional areas and different velocities. The experimental data show that the homogeneity is less than 1.2 %, except at low speeds (e.g. 0.3 m / s), where it is equal to 3.8 %. The impact of the instability cannot be excluded from homogeneity measurements. By performing an additional measurement, we also noted that the blockage effect occurs in the field upstream of the anemometer in the measuring section.

Ključne besede:Velocity measurement, Lasser Doppler anemometer, velocity profile, wind tunnel, homogeneity, stability

Podobna dela

Podobna dela v RUL:
  1. Vpliv uporabe gozdarske osebne varovalne opreme na psihofizične obremenitve
  2. Vpliv uporabe gozdarske osebne varovalne opreme na psihofizične obremenitve
  3. Kardiopulmonalna reanimacija v času izrednih razmer epidemije COVID-19
  4. Negorljivi materiali kot zaščita pri požarih
  5. Ustreznost radiološke osebne varovalne opreme v veterinarskih ambulantah
Podobna dela v drugih slovenskih zbirkah:
  1. Preprečevanje okužb s SARS-Cov-2 pri pacientih, ki se zdravijo s hemodializo
  2. Perioperativna priprava bolnika z okužbo COVID-19
  3. Ozaveščenost medicinskih sester o pravilnem rokovanju z medicinskimi tekstilijami
  4. Pojavnost kožnih sprememb zaradi nošnje osebne varovalne opreme pri zdravstvenih delavcih v času pandemije SARS-CoV-2
  5. Initial study of immersion thermal manikin development and its manufacture from solid blocks

Nazaj