Details

Vpliv kota dotika merjene površine na koordinatnem merilnem stroju
ID Žnidaršič, Lovro (Author), ID Kutin, Jože (Mentor) More about this mentor... This link opens in a new window

.pdfPDF - Presentation file, Download (1,81 MB)
MD5: C305D728619BEEE1AB35FDBF19131777

Abstract
Pri dimenzijskih meritvah na koordinatnem merilnem stroju, imajo že minimalne spremembe v strategiji merjenja lahko znaten vpliv na rezultat meritev. V diplomski nalogi smo obravnavali vpliv naklona tipala na izmerjeno vrednost. Izvedli smo meritve, pri katerih smo se površine kalibracijske krogle v istih točkah dotikali z različnimi nakloni merilnega tipala. Poleg tega smo izmerili še vplive velikosti tipala, pomične hitrosti in premika v skrajno lego. Vse merjene vplive smo po velikosti primerjali tako s predpisanim največjim dopustnim odstopanjem kot tudi med seboj. V splošnem lahko zaključimo, da je vpliv kota tipala zelo velik, vendar ne presega največjega dopustnega merilnega pogreška, ki je opredeljen v specifikacijah merilne naprave.

Language:Slovenian
Keywords:koordinatni merilni stroj, dimenzijsko merjenje, vpliv uporabe, merilni pogrešek, merilno tipalo, naklon tipala
Work type:Bachelor thesis/paper
Typology:2.11 - Undergraduate Thesis
Organization:FS - Faculty of Mechanical Engineering
Place of publishing:Ljubljana
Publisher:[L. Žnidaršič]
Year:2026
Number of pages:XX, 44 str.
PID:20.500.12556/RUL-181204 This link opens in a new window
UDC:53.089.6:621.713.3:621.7.013.2(043.2)
COBISS.SI-ID:273587203 This link opens in a new window
Publication date in RUL:27.03.2026
Views:356
Downloads:180
Metadata:XML DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Secondary language

Language:English
Title:Influence of the angle of contact with the measured surface on coordinate measuring machine
Abstract:
In dimensional measurements on a coordinate measuring machine, even minimal changes in the measurement strategy can have a significant impact on the measurement result. In the thesis, we examined the influence of the probe tilt on the measured value. We performed measurements in which we touched the surface of the calibration sphere at the same points with different measuring probe tilts. In addition, we measured the influences of the probe size, the traversing speed, and the displacement to the extreme position. All measured influences were compared in magnitude both with the prescribed maximum permissible error and among themselves. In general, it can be concluded that the influence of the probe angle is very large, but does not exceed the maximum permissible measuring error that is defined in the specifications of the measuring device.

Keywords:coordinate measuring machine, dimensional measurement, influence of use, measurement error, measuring probe, probe tilt

Similar documents

Similar works from RUL:
Similar works from other Slovenian collections:

Back