Details

Errors in the sampling process when the number of samples in the measurement window is small
ID Agrež, Dušan (Author)

URLURL - Source URL, Visit https://ieeexplore.ieee.org/document/10945338 This link opens in a new window
.pdfPDF - Presentation file, Download (1,63 MB)
MD5: EAF51AB25417152811687442C7942B4F

Abstract
Sampling process is one of the most important part in acquiring information in a measurement system. This article presents an analysis of the sampling of a measured signal that originally takes place in the time space, when the number of samples in the measurement window interval is small. Sampling errors increase under incoherent sampling conditions when the absolute number of samples decreases and even more when the number of samples per signal period decreases. The analysis shows that the number of samples in the measurement interval has the greatest impact on the error in the evaluation of the signal amplitude. This is true if the duty cycle of the sampling pulse or the aperture time is much shorter than the sampling interval, which is a typical case for sampling systems. To determine the maximum sampling errors, incoherent sampling conditions were used with the frequency component positioned at the midpoint of the two frequency bins. The simulation analysis was also confirmed on a real measurement system. The sampling error decreases inversely proportional to the number of samples considering the basic rectangular window over the observation interval. This behavior can be improved by signal processing with stronger windows, such as RVI windows.

Language:English
Keywords:aperture time, number of samples, Rife-Vincent windows class I., sampling process
Work type:Article
Typology:1.01 - Original Scientific Article
Organization:FE - Faculty of Electrical Engineering
Publication status:Published
Publication version:Version of Record
Year:2025
Number of pages:[10] str.
Numbering:Vol. 74, art. 6503510
PID:20.500.12556/RUL-168811 This link opens in a new window
UDC:621.3
ISSN on article:1557-9662
DOI:10.1109/TIM.2025.3555719 This link opens in a new window
COBISS.SI-ID:234032131 This link opens in a new window
Publication date in RUL:25.04.2025
Views:289
Downloads:78
Metadata:XML DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Record is a part of a journal

Title:IEEE transactions on instrumentation and measurement
Shortened title:IEEE trans. instrum. meas.
Publisher:Institute of Electrical and Electronics Engineers
ISSN:1557-9662
COBISS.SI-ID:528591385 This link opens in a new window

Licences

License:CC BY 4.0, Creative Commons Attribution 4.0 International
Link:http://creativecommons.org/licenses/by/4.0/
Description:This is the standard Creative Commons license that gives others maximum freedom to do what they want with the work as long as they credit the author.

Secondary language

Language:Slovenian
Keywords:aperturni čas, število vzorcev, okna Rife-Vincent razreda I., proces vzorčenja

Projects

Funder:ARIS - Slovenian Research and Innovation Agency
Project number:P2-0250
Name:Metrologija in biometrični sistemi