izpis_h1_title_alt

Efficacy of breast shielding during head computed tomography examination
ID Zalokar, Nika (Author), ID Mekiš, Nejc (Author)

.pdfPDF - Presentation file, Download (436,43 KB)
MD5: 9503E592AC2B88C1E033C8C9D86EA95D

Abstract
Background. Female breasts are exposed to scattered radiation regardless of not being included in the primary field during head CT. This study aimed to investigate whether the use of lead shielding is beneficial in dose reduction to the breasts during head CT. Patients and methods. The study was performed in two different hospitals on two different CT units and included 120 patients. Half of the measurements (n = 60) was conducted without the use of lead shielding and the other half (n = 60) with the use of lead shielding of 0.5 mm equivalent thickness. Results. Significant skin dose reduction to the breasts during head CT in both hospitals with the use of lead shielding was discovered; 81% (338.2 ± 43.7 μGy to 64.3 ± 18.8 μGy) in Hospital A and 74% (from 253.1 ± 35.1 μGy to 65.3 ± 16.9 μGy) in Hospital B. Conclusions. Considering the assumed carcinogenic effect of low doses of radiation, high frequency of the head CT scans and the significant reduction of radiation doses to the highly radiosensitive breasts, the use of lead shielding is highly recommendable.

Language:English
Keywords:scatter radiation, head CT, lead shielding, breasts, dose reduction
Typology:1.01 - Original Scientific Article
Organization:ZF - Faculty of Health Sciences
Publication status:Published
Publication version:Version of Record
Year:2021
Number of pages:Str. 116-120
Numbering:Vol. 55, no. 1
PID:20.500.12556/RUL-124556 This link opens in a new window
UDC:616-074
ISSN on article:1318-2099
DOI:10.2478/raon-2020-0044 This link opens in a new window
COBISS.SI-ID:25473795 This link opens in a new window
Publication date in RUL:02.02.2021
Views:806
Downloads:232
Metadata:XML RDF-CHPDL DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Record is a part of a journal

Title:Radiology and oncology
Shortened title:Radiol. oncol.
Publisher:Slovenian Medical Society - Section of Radiology, Croatian Medical Association - Croatian Society of Radiology
ISSN:1318-2099
COBISS.SI-ID:32649472 This link opens in a new window

Licences

License:CC BY 4.0, Creative Commons Attribution 4.0 International
Link:http://creativecommons.org/licenses/by/4.0/
Description:This is the standard Creative Commons license that gives others maximum freedom to do what they want with the work as long as they credit the author.
Licensing start date:02.02.2021

Secondary language

Language:Slovenian
Keywords:razpršeno sevanje, računalniška tomografija glave, svinčeni ščit, prsi, zmanjšanje odmerka

Similar documents

Similar works from RUL:
Similar works from other Slovenian collections:

Back