izpis_h1_title_alt

Določanje frekvenčne prenosne funkcije merilnika tlaka pri dinamičnem umerjanju z udarno cevjo
ID Komel, Žan (Author), ID Kutin, Jože (Mentor) More about this mentor... This link opens in a new window, ID Svete, Andrej (Co-mentor)

.pdfPDF - Presentation file, Download (4,66 MB)
MD5: 3AC1D694EB67DA7C7D336E75079108B7

Abstract
V okviru te naloge smo s pomočjo programskega okolja LabVIEW izdelali program, kateri določi frekvenčno prenosno funkcijo obravnavanega sistema z uporabo hitre Fourierjeve transformacije. Najprej smo jo določevali na simuliranih signalih, kasneje tudi na meritvah, pridobljenih z dinamičnim generatorjem skočne spremembe tlaka, udarno cevjo. Obravnavani so pogreški metode v primerjavi z analitično rešitvijo. Predstavljeni so tudi vplivi lastne frekvence, razmernika dušenja, frekvence vzorčenja in dolžine zajetega signala na točnost metode.

Language:Slovenian
Keywords:frekvenčna prenosna funkcija, merilnik tlaka, udarna cev, dinamično umerjanje, skočni odziv, Fourierjeva transformacija
Work type:Final paper
Typology:2.11 - Undergraduate Thesis
Organization:FS - Faculty of Mechanical Engineering
Place of publishing:Ljubljana
Publisher:[Ž. Komel]
Year:2020
Number of pages:XII, 23 f.
PID:20.500.12556/RUL-119619 This link opens in a new window
UDC:517.443:531.787:004.4(043.2)
COBISS.SI-ID:30670083 This link opens in a new window
Publication date in RUL:10.09.2020
Views:920
Downloads:181
Metadata:XML RDF-CHPDL DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Secondary language

Language:English
Title:Determination of the frequency transfer function of a pressure meter in dynamic calibration with a shock tube
Abstract:
Within this thesis, we developed a computer program that determines the frequency transfer function of the considered system using fast Fourier transform in the LabVIEW programming environment. We first determined it on simulated signals and later also on measured signals obtained with a pressure step generator, the shock tube. Errors of the method in comparison with the analytical solution are considered. Furthermore, the effects of natural frequency, damping ratio, sampling frequency and acquired signal length on the accuracy of the method are also presented.

Keywords:frequency transfer function, pressure meter, shock tube, dynamic calibration, step response, Fourier transform