Ponovljivost meritev anteriornega odmika golenice pri uporabi kolenskega artrometra GNRB
ID Vauhnik, Renata (Author), ID Jeraj, Anamarija (Author), ID Glinšek, Klemen (Author), ID Rugelj, Darja (Author)

.pdfPDF - Presentation file, Download (351,63 KB)
MD5: 8D873961F0DDB72D65470AB8EDF9B70B
URLURL - Source URL, Visit http://www.dlib.si/details/URN:NBN:SI:doc-GIA744MO This link opens in a new window

Abstract
Uvod: Merjenje anteriornega odmika golenice s kolenskim artrometrom je pomemben del kliničnega pregleda kolenskega sklepa. Namen raziskave je bil ugotoviti ponovljivost meritev anteriornega odmika golenice s kolenskim artrometrom GNRB. Metode: Merjenje anteriornega odmika golenice s kolenskim artrometrom je bilo ponovljeno sedemkrat na eni preiskovanki. Med posameznimi ponovitvami je bilo 20 minut premora. Kolenski artrometer GNRB je bil za vsako posamezno meritev ponovno nameščen na spodnji ud preiskovanke. Anteriorni odmik golenice je bil izmerjen pri silah 134 N in 250 N, tako na desnem kot levem kolenu. Za analizo podatkov je bila uporabljena opisna statistika in izračunana je bila standardna napaka meritve. Rezultati: Povprečni anteriorni odmik golenici pri sili 134 N je 7 mm, pri sili 250 N pa 9,9 mm. Standardna napaka meritve se je pri meritvah anteriornega odmika golenice pri sili 134 N in sili 250 N gibala v razponu med 0,152 mm in 0,189 mm. Zaključki: Ponovljivost meritev anteriornega odmika golenice s kolenskim artrometrom GNRB je klinično sprejemljiva, saj je bila standardna napake meritve manjša kot 0,2 mm.

Language:Slovenian
Keywords:anteriorni odmik golenice, laksnost kolenskega sklepa, ocena integritete sprednje križne vezi, dejavnik tveganja, test-retest
Typology:1.01 - Original Scientific Article
Organization:ZF - Faculty of Health Sciences
Year:2014
Number of pages:Str. 27-30
Numbering:Letn. 22, št. 1
PID:20.500.12556/RUL-114742 This link opens in a new window
UDC:615.82/.84:616.728.3
ISSN on article:1318-2102
URN:URN:NBN:SI:doc-GIA744MO
COBISS.SI-ID:4693355 This link opens in a new window
Publication date in RUL:06.03.2020
Views:1746
Downloads:204
Metadata:XML RDF-CHPDL DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Record is a part of a journal

Title:Fizioterapija
Shortened title:Fizioterapija
Publisher:Društvo fizioterapevtov Slovenije, Društvo fizioterapevtov Slovenije, Združenje fizioterapevtov Slovenije
ISSN:1318-2102
COBISS.SI-ID:30074880 This link opens in a new window

Secondary language

Language:English
Title:Repeatability of knee anterior laxity assessment when using GNRB knee arthrometer
Abstract:
Background: Assessment of anterior tibial displacement is the crucial part of clinical examination of the knee joint. The purpose of this study was to investigate repeatability of GNRB knee arthrometer. Methods: Measurements of knee anterior laxity were repeated seven times on one subject with 20 minutes between each repetition. Knee arthrometer GNRB was removed from the subject’s knee after each measurement. Knee anterior laxity was measured at force 134N and 250N on the right and left knee, respectively. Data was analysed with descriptive statistics. Standard error of measurement (SEM) was calculated in order to define the stability of repeated measurements. Results: Average knee anterior laxity is 7 mm at force 134N, while at force 250N, average knee anterior laxity is 9.9 mm. Standard error of measurement was ranging from 0.152 to 0.189 mm at both forces 134N and 250N, respectively. Conclusions: Repeatability of knee anterior laxity measured with GNRB knee arthrometer is clinically acceptable, since the SEM was lower than 0.2 mm.

Keywords:anterior tibial displacement, knee laxity, assessment of anterior cruciate integrity, risk factor, test-retest