Abstract
Uvod: Kontrola kakovosti negatoskopov zajema meritve svetilnosti in homogenosti svetilnosti negatoskopov ter osvetljenost prostora, kjer se negatoskop nahaja. WHO (1992), CEC (1997, cit. po McCarthy in Brennan, 2003), NORDIC (1999) in IPEM (2005) priporočila določajo meje sprejemljivosti zgoraj omenjenih parametrov. Namen: Namen članka je bil izvesti meritve navedenih parametrov in oceniti, v kolikšni meri lastnosti negatoskopov v treh izbranih slovenskih bolnišnicah ustrezajo mednarodnim priporočilom. Metode dela: V raziskavo so bile vključene tri večje slovenske bolnišnice, parametri so bili pridobljeni na skupno 37 negatoskopih na deloviščih radioloških inženirjev. Za izvedbo meritev svetilnosti negatoskopov in osvetljenosti prostorov je bil uporabljen ustrezen umerjen merilnik Light-O-Meter. Za izračun homogenosti svetilnosti se uporabi ustrezna formula. Rezultati in razprava: Minimalne standarde povprečne svetilnosti dosega 51% negatoskopov, homogenosti svetilnosti 72%, osvetljenost prostorov pa je primerna v 76%, kjer se nahajajo negatoskopi. Sklep: Kljub boljšim rezultatom v primerjavi s sorodnimi raziskavami, še vedno obstajajo možnosti za izboljšave. Zato so predlagani ukrepi za izboljšanje delovanja negatoskopov. V obseg programa kontrole kvalitete na radiološki diagnostiki bi bilo potrebno vključiti tudi kontrolo kvalitete negatoskopov.

Language:Slovenian
Keywords:negatoskop, svetilnost negatoskopa, homogenost svetilnosti negatoskopa, osvetljenost prostora
Typology:1.04 - Professional Article
Organization:ZF - Faculty of Health Sciences
Year:2010
Number of pages:Str. 4-10
Numbering:Letn. 27, št. 1
PID:20.500.12556/RUL-114441 This link opens in a new window
UDC:616-07
ISSN on article:1855-5136
URN:URN:NBN:SI:DOC-GUP1JGZK
COBISS.SI-ID:27876313 This link opens in a new window
Publication date in RUL:28.02.2020
Views:580
Downloads:88
Metadata:XML RDF-CHPDL DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Record is a part of a journal

Title:Bilten
Publisher:Društvo radioloških inženirjev Slovenije, Zbornica radioloških inženirjev Slovenije
ISSN:1855-5136
COBISS.SI-ID:241400320 This link opens in a new window

Secondary language

Language:English
Title:Quality control of viewing boxes
Abstract:
Introduction: Quality control of viewing boxes consists of measurements of viewing box brightness, viewing box brightness uniformity and ambient lighting levels in radiographer viewing areas. WHO (1992), CEC (1997, according to McCarthy in Brennan, 2003), NORDIC (1999) and IPEM (2005) recommendations determine the acceptable levels of the above mentioned parameters. Purpose: The purpose of this article was to measure the brightness and brightness uniformity of viewing boxes and ambient lighting levels in three Slovenian hospitals and to compare the results with the international recommendations. Methods: Three major Slovenian hospitals were included in this study. The parameters were acquired on 37 viewing boxes in radiographer viewing areas. An appropriate calibrated meter Light-O-Meter was used for the measurements and a recommended equation was used to calculate brightness uniformity. Results and discussion: 51 % of the viewing boxes met the minimum standards of the average viewing box brightness, 72 % of the viewing boxes met the minimum standards for brightness uniformity and 76 % of radiographer viewing areas met the criteria for the ambient lighting level. Conclusion: This study showed that the measured parameters of viewing boxes included in this article were better than the results from other studies and for that the measures to improve the functioning of viewing boxes were suggested. It is necessary to include the quality control of viewing boxes in the quality control programme at the diagnostic radiology department.