V diplomskem delu so opisani preskusi, ki so aplikabilni za operacijsko svetilko tako po medicinskem osnovnem standardu IEC/EN 60601-1:2006+A11:2011+A1:2013, kot tudi dodatnemu IEC/EN 60601-2-41:2010. Na začetku predstavim, zakaj potrebujemo standard in vsakega posebej opišem. V nadaljevanju je opisana črna soba oziroma fotobiološki laboratorij. Ta del nam pove, zakaj je potrebno izvajati preskuse v takem okolju in kako ljudje zaznavamo svetlobo. Spoznamo tudi instrumente in pripomočke, katere sem potreboval za izvedbo meritev. Sledi opis operacijske svetilke in njenih lastnosti. Nato se začnejo opisi in razlage preskušanj, ki sem jih opravljal v fotobiološkem laboratoriju. Preskusi si sledijo po vrsti, kot je to zapisano v standardu. Od meritev sredinske osvetlitve, premera svetlobnega polja, vpliva senc, globine osvetlitve do ostalih preskusov, ki se navezujejo na barvno temperaturo, CRI indeks, UV svetlobo, barvni prostor. Po končanih fotobioloških preskusih pa sem se osredotočil na električne preskuse, ki jih zahteva osnovni medicinski standard. Tu govorim o več različnih preskusih, za vsakega razložim, zakaj ga izvajamo in kako so potekale meritve. To so meritve porabe, temperaturni preskusi, prebojna trdnost in še nekaj ostalih. Omenim tudi določene mehanske preskuse, ki jih je bilo potrebno izvesti. Za zaključek pa razložim, kaj sem se naučil in predstavim možne izboljšave za natančnejše meritve osvetljenosti.
Language:
Slovenian
Keywords:
operacijska svetilka
,
preskus
,
medicinski standard
Work type:
Bachelor thesis/paper
Organization:
FE - Faculty of Electrical Engineering
Year:
2019
PID:
20.500.12556/RUL-108669
Publication date in RUL:
11.07.2019
Views:
1357
Downloads:
255
Metadata:
Cite this work
Plain text
BibTeX
EndNote XML
EndNote/Refer
RIS
ABNT
ACM Ref
AMA
APA
Chicago 17th Author-Date
Harvard
IEEE
ISO 690
MLA
Vancouver
:
KRAŠEVEC, URBAN, 2019,
Preizkušanje delovanja operacijske luči
[online]. Bachelor’s thesis. [Accessed 26 April 2025]. Retrieved from: https://repozitorij.uni-lj.si/IzpisGradiva.php?lang=eng&id=108669
Copy citation
Share:
Secondary language
Language:
English
Title:
Testing of the surgical light
Abstract:
This diploma thesis presents description of test procedures which are applicable for surgical luminaries according to the basic medical standard IEC/EN 60601-1:2006+A11:2011+A1:2013 and additional standard IEC/EN 60601-2-41:2010. Initially, I introduce basic needs for standardization and describe each standard one by one. The description of a black room also known as photobiological laboratory is followed next. This part talks about why tests have to be performed at a specific environment and how people perceive light. Instruments and other accessories which are used for measurements are described. Then we continue with a description of an operation lamp and its properties. The main part of this thesis are descriptions and explanations of tests which have been done in photobiological laboratory. Tests are performed in sequence according to the standard. From measuring central illuminance, light field diameter, shadow dilution, depth of illumination and other test procedures which are related to color temperature known as CRI index, UV light and color space. After photobiological test have been performed, the main focus was on electrical tests which are required by medical standard. Each test is explained and the reason why it was performed, for example input test, temperature test, dielectric strength test, mechanical tests and few others. Finally possible improvements for more accurate measurements of illuminance are described.
Keywords:
operational lamp
,
test procedure
,
medical standard
Similar documents
Similar works from RUL:
The integration of CRM systems
Distributed representation based classification
Algorithm as a service
QRS complex detection of electrocardiogram
Interactive reading of rainfall graphs
Similar works from other Slovenian collections:
Uporaba orodij za prognoziranje daljinskega prometa, primer Trans-Tools
Spremembe intenzivnosti erozije v porečju Dragonje v drugi polovici 20. stoletja
Prihodnji scenariji rabe tal na celotnem območju Parka Škocjanske jame z uporabo prostorskega modela DynaCLUE
Grassland ecology under changing climate and land use
Back