izpis_h1_title_alt

The importance of additional geometrical verification with auto beam hold function in prostate cancer irradiation
ID Salmič, Klemen (Avtor), ID Matjašič, Alenka (Avtor), ID Žager Marciuš, Valerija (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (777,91 KB)
MD5: 7C27DCBD0D91C839FDB53B1793FB7CC9
.pdfPDF - Predstavitvena datoteka, prenos (791,02 KB)
MD5: 337931057E01F31E8D3A451B3043E700

Izvleček
This paper aims to determine the need for additional imaging verification when using the auto beam hold function in radiotherapy of prostate cancer patients with inserted gold fiducial markers. Forty patients who underwent irradiation of the prostate and twenty with prostate and pelvic lymph node radiotherapy were included in the retrospective study. Intrafraction shifts during irradiation were compared with the auto beam hold function in the translational directions. The function was used with time tracking (5 seconds), a tolerance limit for marker deviations (4 mm), and bone structures in the large irradiation field (5 mm). The need for additional image verification was higher in the larger irradiation field group. When translational shifts were analysed, a statistically significant difference in the vertical direction was found in the group with only prostate irradiation (p = 0.013). A statistically significant difference in the lateral direction was found in the group with a larger irradiation field (p = 0.021). Translational shifts were not statistically significantly different between the two groups (p > 0.05). Conclusion: Intrafraction shifts of the prostate increase the need for additional imaging verification. The use of the auto beam hold function is effective in reducing errors.

Jezik:Angleški jezik
Ključne besede:prostate cancer, radiotherapy, intrafraction movement, 2-D kV imaging in radiotherapy, auto beam hold function
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:ZF - Zdravstvena fakulteta
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Leto izida:2023
Št. strani:Str. 301-306
Številčenje:Vol. 38, no. 4
PID:20.500.12556/RUL-156054 Povezava se odpre v novem oknu
UDK:616-073:616-006
ISSN pri članku:1451-3994
DOI:10.2298/NTRP2304301S Povezava se odpre v novem oknu
COBISS.SI-ID:194385667 Povezava se odpre v novem oknu
Datum objave v RUL:06.05.2024
Število ogledov:421
Število prenosov:53
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:Nuclear Technology & Radiation Protection
Skrajšan naslov:Nucl. Technol. Radiat. Prot.
Založnik:Vinča Instituta of Nuclear Sciences
ISSN:1451-3994
COBISS.SI-ID:18296103 Povezava se odpre v novem oknu

Licence

Licenca:CC BY-NC-ND 3.0, Creative Commons Priznanje avtorstva-Nekomercialno-Brez predelav 3.0 Nedoločena
Povezava:https://creativecommons.org/licenses/by-nc-nd/3.0/deed.sl
Opis:Dovoljuje kopiranje in razširjanje vsebin v kakršnemkoli mediju in obliki. Primerno morate navesti avtorja, povezavo do licence in označiti spremembe, če so kakšne nastale. To lahko storite na kakršenkoli razumen način, vendar ne na način, ki bi namigoval na to, da dajalec licence podpira vas ali vašo uporabo dela. Te vsebine ne smete uporabiti v komercialne namene. Če vsebino predelate (remixate), uredite ali na njej gradite, spremenjene vsebine ne smete razširjati. Ne smete uporabiti pravnih določil ali tehničnih ukrepov, ki bi pravno omejili ali onemogočilo druge, da bi storili karkoli, kar licenca dovoli.

Podobna dela

Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:

Nazaj