izpis_h1_title_alt

Verification, implementation and harmonization of automated chemiluminescent immunoassays for MPO- and PR3-ANCA detection
ID Ogrič, Manca (Avtor), ID Švec, Tinka (Avtor), ID Mrak Poljšak, Katjuša (Avtor), ID Žigon, Polona (Avtor), ID Hočevar, Alojzija (Avtor), ID Čučnik, Saša (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (514,85 KB)
MD5: E6E9F0F9C540CDE90EF8D1FF1CF177C2
URLURL - Izvorni URL, za dostop obiščite https://www.degruyter.com/document/doi/10.1515/cclm-2023-0764/html Povezava se odpre v novem oknu

Izvleček
Objectives: Antineutrophil cytoplasmic antibody (ANCA) testing assists clinicians diagnose ANCA-associated vasculitis (AAV). We aimed to verify and harmonize chemiluminescent immunoassays for the detection of myeloperoxidase (MPO)- and proteinase 3 (PR3)-ANCA. Methods: An in-house ELISA, a capture ELISA, and a chemiluminescent assay QUANTA Flash on a BIO-FLASH analyzer were used to detect MPO- and PR3-ANCA in sera from 39 patients with AAV, 55 patients with various non-AAV, and 66 patients with connective tissue diseases. The results of the assays were evaluated, and their clinical performance was assessed. The precision and linearity of the QUANTA Flash assays were determined, and likelihood ratios (LRs) for AAV at diagnosis were calculated. Results: The precision and linearity of the QUANTA Flash assays were confirmed. Overall agreement between 97.5 and 98.8 % and Cohen’s kappa coefficients between 0.861 and 0.947 were observed for the results of the QUANTA Flash assays and ELISAs. The diagnostic sensitivity, specificity, and ROC analysis of the assays for AAV were statistically similar (in-house ELISA 89.7 %, 95.0 %, and 0.937; capture ELISA 92.3 %, 98.3 %, and 0.939; and QUANTA Flash 89.7 %, 95.9 %, and 0.972). For the QUANTA Flash assay results, the interval-specific LRs for AAV at diagnosis were: 0–8 CU had LR 0.08, 8–29 CU had LR 1.03, 29–121 CU had LR 7.76, 121–191 CU had LR 12.4, and >191 CU had LR ∞. Conclusions: The QUANTA Flash MPO and PR3 assays provide precise and consistent results and have comparable clinical utility for AAV. The calculated LRs were consistent with published LRs, confirming the utility of LRs for harmonization of ANCA results.

Jezik:Angleški jezik
Ključne besede:antineutrophil cytoplasmic antibodies, ANCA, proteinase 3, PR3, myeloperoxidase, MPO, ANCA-associated vasculitis, AAV, chemiluminescent immunoassay, likelihood ratios, LRs
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:MF - Medicinska fakulteta
FFA - Fakulteta za farmacijo
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Leto izida:2024
Št. strani:Str. 682–689
Številčenje:Vol. 62, iss. 4
PID:20.500.12556/RUL-154819 Povezava se odpre v novem oknu
UDK:616-097
ISSN pri članku:1437-4331
DOI:10.1515/cclm-2023-0764 Povezava se odpre v novem oknu
COBISS.SI-ID:169691907 Povezava se odpre v novem oknu
Datum objave v RUL:04.03.2024
Število ogledov:424
Število prenosov:44
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:Clinical chemistry and laboratory medicine
Skrajšan naslov:Clin. chem. lab. med.
Založnik:De Gruyter, European Federation of Clinical Chemistry and Laboratory Medicine
ISSN:1437-4331
COBISS.SI-ID:680355 Povezava se odpre v novem oknu

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:antinevtrofilna citoplazemska protitelesa, ANCA, proteinaza 3, PR3, mieloperoksidaza, MPO, vaskulitis, povezan z ANCA, AAV, kemiluminiscentni imunski test, razmerja verjetnosti, LR

Projekti

Financer:ARRS - Agencija za raziskovalno dejavnost Republike Slovenije

Podobna dela

Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:

Nazaj