Gridless adult cervical spine radiography and its' effect on image quality and radiation dose : a phantom study
ID
Mekiš, Nejc
(
Avtor
),
ID
Bianchi, Tiziano Alessandro
(
Avtor
),
ID
Doyle, Clodagh
(
Avtor
),
ID
Gauchat, Manon Pauline
(
Avtor
),
ID
Geerling, Indy Larissa Amata
(
Avtor
),
ID
Linneman, Jesse
(
Avtor
),
ID
Staats, Seugnet
(
Avtor
),
ID
Campeanu, Cosmin
(
Avtor
)
PDF - Predstavitvena datoteka,
prenos
(867,91 KB)
MD5: EE596952F5BBBFCAF0CE6DE600695039
URL - Izvorni URL, za dostop obiščite
https://www.sciencedirect.com/science/article/pii/S1078817423002493
Galerija slik
Izvleček
Introduction: Imaging of the cervical spine in general radiography is most frequently performed using an anti-scatter grid. The purpose of this study was to investigate the effects of a gridless setting on image quality and radiation dose during digital radiography of the anteroposterior (AP) and lateral (LAT) cervical spine. Methods: A phantom study was performed with a variety of tube voltages (63e75 kV) with and without an anti-scatter grid. The tube current time product (mAs) and dose area product (DAP) were recorded and used to calculate effective dose (ED) and individual organ dose using PCXMC 2.0 software, as well as entrance surface dose (ESD) and objective image quality: signal-to-noise ratio (SNR) and contrast-tonoise ratio (CNR). Subjective visual image quality grading characteristics (VGC) was performed by five qualified radiographers. Results: In a gridless setting, the AP and LAT positions showed significantly lower DAP (1.6 mGym2 ; 61.3 % and 1.6 mGym2 ; 51.2%), ESD (27.6 mGy; 57.3% and 77.2 mGy; 47.2%) and ED (4.2 mSv; 61.3% and 2.3 mSv; 48.9%). In a gridless setting in the AP position, there is a slight significant deterioration in image quality. In the lateral projection, on the other hand, the image quality without the use of grid was only significantly reduced in three of six criteria and there was no difference in the objective image quality between the two settings examined. Conclusion: The results of this study show that gridless setting significantly decreases radiation dose and image quality, but the quality in the lateral projection is still acceptable for diagnostic purpose. Implications for practice: The protocol without the use of the anti-scatter grid in cervical spine radiography leads to a reduction in the radiation dose in both projections, but the image quality in the AP is significantly reduced for all criteria examined, with a slight deterioration in image quality in the lateral projection.
Jezik:
Angleški jezik
Ključne besede:
antero-posterior C-spine
,
lateral C-spine
,
anti-scatter grid
,
digital radiography
,
image quality
,
dose reduction
Vrsta gradiva:
Članek v reviji
Tipologija:
1.01 - Izvirni znanstveni članek
Organizacija:
ZF - Zdravstvena fakulteta
Status publikacije:
Objavljeno
Različica publikacije:
Objavljena publikacija
Datum objave:
22.12.2024
Leto izida:
2024
Št. strani:
Str. 359-366
Številčenje:
Vol. 30, iss. 1
PID:
20.500.12556/RUL-153407
UDK:
616-073:779
ISSN pri članku:
1078-8174
DOI:
10.1016/j.radi.2023.12.009
COBISS.SI-ID:
179374083
Datum objave v RUL:
04.01.2024
Število ogledov:
610
Število prenosov:
57
Metapodatki:
Citiraj gradivo
Navadno besedilo
BibTeX
EndNote XML
EndNote/Refer
RIS
ABNT
ACM Ref
AMA
APA
Chicago 17th Author-Date
Harvard
IEEE
ISO 690
MLA
Vancouver
:
Kopiraj citat
Objavi na:
Gradivo je del revije
Naslov:
Radiography
Skrajšan naslov:
Radiography
Založnik:
Elsevier, The College of Radiographers
ISSN:
1078-8174
COBISS.SI-ID:
383851
Licence
Licenca:
CC BY-NC-ND 4.0, Creative Commons Priznanje avtorstva-Nekomercialno-Brez predelav 4.0 Mednarodna
Povezava:
http://creativecommons.org/licenses/by-nc-nd/4.0/deed.sl
Opis:
Najbolj omejujoča licenca Creative Commons. Uporabniki lahko prenesejo in delijo delo v nekomercialne namene in ga ne smejo uporabiti za nobene druge namene.
Sekundarni jezik
Jezik:
Slovenski jezik
Ključne besede:
antero-posteriorna C-hrbtenica
,
lateralna C-hrbtenica
,
mreža proti sipanju
,
digitalna radiografija
,
kakovost slike
,
zmanjšanje doze
Podobna dela
Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:
Nazaj