izpis_h1_title_alt

60 years of Charnley–Muller Alivium hip prosthesis : the revision percentage and tribo-corrosion sequelae after a mean of 27 years
ID Milošev, Ingrid (Author), ID Trebše, Rihard (Author), ID Cör, Andrej (Author), ID Levašič, Vesna (Author)

.pdfPDF - Presentation file, Download (1,96 MB)
MD5: FD681CEBCAD7D7835C4EF2F29610B432
URLURL - Source URL, Visit https://link.springer.com/article/10.1007/s00402-023-04824-y This link opens in a new window

Abstract
Introduction The main aim was to analyse the series of 29 collected cemented Charnley–Muller Alivium retrievals with the meantime in situ of 27 years. In addition, the revision rate of 1425 Alivium prostheses implanted at our institution between 1977 and 1992 was calculated. Materials and methods The revision percentage of the Alivium cohort was calculated up to 45 years of follow-up and compared to that of all total hip arthroplasties (THAs) implanted in the same period (No. 5535). Metal and polyethylene retrieved components were inspected in 29 cases for wear damage and roughness. Wear particles were retrieved from periprosthetic tissue using digestion protocols and their composition, morphology, and size distribution were investigated. Periprosthetic tissue was analysed histologically. Results The revision percentage of the Alivium cohort was 16% at 45 years of follow-up. It was comparable to all the THAs implanted at the same time (18%). The shape of polyethylene particles isolated from periprosthetic tissue corresponded to the wear pattern on polyethylene cups. Polyethylene particles were the main wear product, with the majority (68%) of particles smaller than 0.1 µm. Metal particles were rare with two types: CoCr and Cr based. Histological analysis showed that in 14 out of 18 specimens, the metal particles were graded + 1, reflecting that the metal loading in the periprosthetic tissue was low. Conclusions Our study represents valuable data not reported previously on the survival rate of Charnley–Muller prostheses at 45 years of follow-up and a unique insight into the collected retrievals from the materials’ point of view.

Language:English
Keywords:Charnley–Muller hip prosthesis, wear, wear particles, revision percentage
Work type:Article
Typology:1.01 - Original Scientific Article
Organization:MF - Faculty of Medicine
Publication status:Published
Publication version:Version of Record
Year:2023
Number of pages:Str. 6021-6031
Numbering:Vol. 143, iss. 10
PID:20.500.12556/RUL-151922 This link opens in a new window
UDC:617.3
ISSN on article:1434-3916
DOI:10.1007/s00402-023-04824-y This link opens in a new window
COBISS.SI-ID:146148611 This link opens in a new window
Publication date in RUL:25.10.2023
Views:399
Downloads:34
Metadata:XML DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Record is a part of a journal

Title:Archives of orthopaedic and trauma surgery
Shortened title:Arch. orthop. trauma surg.
Publisher:Springer Nature
ISSN:1434-3916
COBISS.SI-ID:513633817 This link opens in a new window

Licences

License:CC BY 4.0, Creative Commons Attribution 4.0 International
Link:http://creativecommons.org/licenses/by/4.0/
Description:This is the standard Creative Commons license that gives others maximum freedom to do what they want with the work as long as they credit the author.

Secondary language

Language:Slovenian
Keywords:kolčna proteza Charnley-Muller, obraba, obrabni delci, odstotek revizije