izpis_h1_title_alt

Izboljšava vodenja vrat stroja namenjena pranju in dezinfekciji medicinskih pripomočkov
ID Rogar, Domen (Author), ID Klemenc, Jernej (Mentor) More about this mentor... This link opens in a new window

.pdfPDF - Presentation file, Download (5,24 MB)
MD5: AD7E27D85AF1FDE3098ACBC40D5FC91D
.pdfPDF - Appendix, Download (58,96 KB)
MD5: D4083FE281F0CC8F7D326C30E25AD93B
.pdfPDF - Appendix, Download (60,22 KB)
MD5: 9EB08DF6EA932C7E723037495B032F41

Abstract
V magistrskem delu je bilo treba izboljšati vodenje vrat stroja, namenjenega pranju in dezinfekciji medicinskih pripomočkov. Ob odpiranju vrat pralnika ta udarijo v steno komore pralnika, kar povzroči zlom navojne palice, na katero so obešena vrata. Pregled delovanja mehanizma, analiza poškodbe navojne palice in izdelava kinematike gibanja vrat so nam omogočili izdelavo različnih konceptov preprečitve loma navojne palice. Koncepte smo nato vrednotili po tehničnem in ekonomskem kriteriju ter izbrali najučinkovitejši koncept. Izbran koncept smo nato detajlno zmodelirali in pripravili tehnično dokumentacijo. Izvedena je bila tudi trdnostna analiza kritičnih komponent koncepta.

Language:Slovenian
Keywords:pralnik in dezinfektor, kinematika gibanja vrat, analiza poškodbe, konstruiranje, metodika konstruiranja, tehnična dokumentacija, trdnostna analiza
Work type:Master's thesis/paper
Typology:2.09 - Master's Thesis
Organization:FS - Faculty of Mechanical Engineering
Place of publishing:Ljubljana
Publisher:[]D. Rogar
Year:2022
Number of pages:XX, [78] str.
PID:20.500.12556/RUL-143545 This link opens in a new window
UDC:004.925.84:519.8+531.1(043.2)
COBISS.SI-ID:140210947 This link opens in a new window
Publication date in RUL:24.12.2022
Views:688
Downloads:147
Metadata:XML DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Secondary language

Language:English
Title:Improvement of machine door guidance for washing and disinfection of medical devices
Abstract:
The objective of the master's degree was to improve the door guidance of machine intended for washing and disinfecting medical devices. When a door opens it hits the chamber wall, causing the screw rod on which the door is hung on to break. Background overview of the door opening mechanism, broken rod analysis, and creation of door opening kinematics allowed us to create different concepts to prevent the breakage of the threaded rod. These concepts were then evaluated according to the technical and financial criteria. After the concept evaluation phase, the best concept according to our criteria was chosen. For the chosen concept we then prepared a detailed model, technical documentation, and stress-strain analysis for critical components.

Keywords:washing machine and disinfectant, door kinematics, damage analysis, design, design methodology, technical documentation, stress-strain analysis

Similar documents

Similar works from RUL:
Similar works from other Slovenian collections:

Back