izpis_h1_title_alt

Evaluation of the size-of-source effect in thermal imaging cameras
ID Pušnik, Igor (Author), ID Geršak, Gregor (Author)

.pdfPDF - Presentation file, Download (8,31 MB)
MD5: 35E544CD2E6633C4C57AA4AB94AFF29A
URLURL - Source URL, Visit https://www.mdpi.com/1424-8220/21/2/607 This link opens in a new window

Abstract
In numerous applications, including current body temperature monitoring in viral pandemic management, thermal imaging cameras are used for quantitative measurements. These require determination of the measurement accuracy (error) and its traceability (measurement uncertainty). Within error estimation, the size-of-source effect (SSE) is an important error source. The SSE is the relation between the physical size of a target and the instrument’s nominal target size. This study presents a direct evaluation of the error due to the SSE. A stable and uniform temperature, generated by blackbodies, was measured by a high-quality thermal imager. To limit the generated radiation, custom-made blocking tiles with different apertures were used. Effects of aperture shapes and positions, camera-target distances and temperature levels on the error were investigated. The study findings suggest that due to the SSE the measured temperatures are too low, especially at longer camera-target distances. The SSE error depends on the number of pixels available and included into the region of interest, for which the accurate measurement is about to be performed. For an accurate temperature measurement, an array of at least 10 × 10 pixels should be exposed to the observed target radiation, while 3 × 3 central pixel area should be included in the temperature calculation.

Language:English
Keywords:temperature, thermal imaging camera, accuracy, size-of-source effect, SSE, region of interest, ROI
Work type:Article
Typology:1.01 - Original Scientific Article
Organization:FE - Faculty of Electrical Engineering
Publication status:Published
Publication version:Version of Record
Year:2021
Number of pages:22 str.
Numbering:Vol. 21, iss. 2, art. 607
PID:20.500.12556/RUL-134901 This link opens in a new window
UDC:621.317
ISSN on article:1424-8220
DOI:10.3390/s21020607 This link opens in a new window
COBISS.SI-ID:47911683 This link opens in a new window
Publication date in RUL:10.02.2022
Views:1082
Downloads:161
Metadata:XML DC-XML DC-RDF
:
Copy citation
Share:Bookmark and Share

Record is a part of a journal

Title:Sensors
Shortened title:Sensors
Publisher:MDPI
ISSN:1424-8220
COBISS.SI-ID:10176278 This link opens in a new window

Licences

License:CC BY 4.0, Creative Commons Attribution 4.0 International
Link:http://creativecommons.org/licenses/by/4.0/
Description:This is the standard Creative Commons license that gives others maximum freedom to do what they want with the work as long as they credit the author.
Licensing start date:16.01.2021

Secondary language

Language:Slovenian
Keywords:temperatura, termovizijske kamere, točnost, učinek velikosti vira, SSE, področje zanimanja, ROI

Projects

Funder:ARRS - Slovenian Research Agency
Project number:P2-0225
Name:Metrologija in kakovost

Funder:Other - Other funder or multiple funders
Funding programme:Republic of Slovenia, Ministry of Economic Development and Technology, Metrology Institute
Project number:6401-18/2008/70

Similar documents

Similar works from RUL:
Similar works from other Slovenian collections:

Back