<metadata xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:dc="http://purl.org/dc/elements/1.1/"><dc:title>Virtual instruments for flow-based and discrete analytical methods</dc:title><dc:creator>Kuljanin,	Aleksandra	(Avtor)
	</dc:creator><dc:creator>Gros,	Nataša	(Mentor)
	</dc:creator><dc:creator>Lotrič,	Uroš	(Komentor)
	</dc:creator><dc:subject>LabVIEW</dc:subject><dc:subject>virtual instruments</dc:subject><dc:subject>multifunctional analytical platforms</dc:subject><dc:subject>multipoint detection</dc:subject><dc:subject>conductivity</dc:subject><dc:subject>sequential injection analysis</dc:subject><dc:description>In the field of analytical chemistry, the use of virtual instruments is still mostly limited 
to signal acquisition and instrument control. The developed platforms are focused on a specific 
analytical method or determination. By utilizing the multichannel capabilities of the LabVIEW 
environment, platforms of wider functionality can be created, and there are many possibilities
for further development and improvements.
The starting point was an already developed, virtually-supported LabVIEW platform for flow-based and discrete analytical methods. The work within this thesis was focused on additional 
extension and upgrading of the platform. Through software modifications, hardware 
components for multipoint conductometric detection were integrated into the virtually-supported sequential injection analysis system. Furthermore, a new virtual instrument was 
developed to process peak signals and calculate key zone penetration parameters. The 
spectrometric microchamber, part of the virtually-supported microtitrator, was redesigned for 
conductometric monitoring.
The number of physical and chemical factors affecting the SIA-gram peak profile is large, and 
the behavior of the plug within the liquid conduits cannot be accurately predicted. Developed 
virtual instrument for peak processing and calculation of most important zone penetration 
parameters in combination with multipoint conductivity monitoring enabled prompt and 
accurate insight into the behavior of the plugs of different chemical composition. The influences
of flow rate, plug volume and coil length on peak profiles and peak overlapping, for plugs of 
different chemical composition, were studied by using multivariate analysis. Under studied set 
of conditions, it was shown that the total overlapping area parameter highly correlate with 
experimental peak absorbance maxima values (R2 &gt; 0.99) for the reactions that employ the 
same plug medium. This was confirmed on the example of two reactions with different kinetics, 
the reaction of iron(II) with ferrozine and the reaction of copper(II) with EDTA. By applying 
the established correlation, absorbance peak maxima can be calculated for the conditions which 
were not tested experimentally, which significantly reduces the total number of experiments 
and chemical consumption in later optimization phases. It was also shown that detailed studies 
of the total overlapping area and the overlapping fraction of the reagent peak can be useful for 
defining the parameters’ range for high reagent utilization and adequate method sensitivity. 
The operating capabilities of the redesigned microchamber were confirmed for monitoring urea 
decomposition catalysed by Jack bean urease. In this way, the platform gained additional 
functionality as another detection method was integrated into the virtually-supported 
microtitrator.
The virtually-supported multipoint conductometric detection is a starting point for the 
introduction of other electroanalytical methods and further extension of the platform. The 
spectrometric microchamber can be remodeled into a dual-detection microchamber, to 
simultaneously monitor changes in absorbance and conductivity.</dc:description><dc:date>2023</dc:date><dc:date>2023-04-17 15:15:02</dc:date><dc:type>Doktorsko delo/naloga</dc:type><dc:identifier>145324</dc:identifier><dc:identifier>VisID: 7641</dc:identifier><dc:identifier>COBISS_ID: 149992707</dc:identifier><dc:language>sl</dc:language></metadata>