Podrobno

Evaluation of simplified ROI definition in facial thermography for psychophysiological and medical applications
ID Stanić, Valentina (Avtor), ID Geršak, Gregor (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (2,37 MB)
MD5: 45E7FBD4A42ADD6D7B247FBFEB962C02
URLURL - Izvorni URL, za dostop obiščite https://www.mdpi.com/2076-3417/16/10/4631 Povezava se odpre v novem oknu

Izvleček
Consistent definition of facial regions of interest (ROIs) is essential for comparable evaluation of temperature dynamics, yet manual ROI annotation remains common. As anatomically faithful annotation is time-consuming and cognitively demanding, this study investigates whether simplified geometric ROIs anchored to facial landmarks can be used without introducing differences in temperature statistics that exceed relevant measurement uncertainty levels. Three novice raters annotated five common facial ROIs across repeated sessions. Temperature metrics (mean, median, and standard deviation) were calculated from all pixels and from the warmest 10% of pixels at three spatial resolutions. Differences between annotation approaches were evaluated with respect to the best achievable and realistic uncertainty levels, 0.3 °C and 1.3 °C, respectively. The simplification reduced annotation time by 42% while maintaining excellent inter- and intra-rater reliability. For most ROIs and metrics, differences remained within the stricter uncertainty level, with larger deviations observed mainly in the anatomically variable periorbital region. Lower spatial resolution increased variability, particularly for small ROIs that fell below the minimum size criterion. These results demonstrate that simplified ROI definitions provide comparable temperature statistics while substantially improving time efficiency, supporting their use in manual and automated facial thermography analysis.

Jezik:Angleški jezik
Ključne besede:infrared thermography, region of interest, facial thermography, thermal image analysis, ROI definition, measurement uncertainty
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:FE - Fakulteta za elektrotehniko
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Leto izida:2026
Št. strani:12 str.
Številčenje:Vol. 16, issue 10, art. 4631
PID:20.500.12556/RUL-182421 Povezava se odpre v novem oknu
UDK:621.317:159.91:61
ISSN pri članku:2076-3417
DOI:10.3390/app16104631 Povezava se odpre v novem oknu
COBISS.SI-ID:277685763 Povezava se odpre v novem oknu
Datum objave v RUL:11.05.2026
Število ogledov:209
Število prenosov:199
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:Applied sciences
Skrajšan naslov:Appl. sci.
Založnik:MDPI
ISSN:2076-3417
COBISS.SI-ID:522979353 Povezava se odpre v novem oknu

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:infrardeča termovizija, območje zanimanja, obrazna termografija, analiza termogramov, definicija ROI, merilna negotovost

Projekti

Financer:ARIS - Javna agencija za znanstvenoraziskovalno in inovacijsko dejavnost Republike Slovenije
Številka projekta:P2-0225
Naslov:Metrologija in kakovost

Podobna dela

Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:

Nazaj