Evaluation of radiation integrity and transmission in lead-rubber layers : a comparative analysis of visual-palpatory, light source and radiographic inspection methods
ID Likar, Manca (Avtor), ID Dolenc, Laura (Avtor), ID Jurša, Laura (Avtor), ID Mekiš, Nejc (Avtor)

.pdfPDF - Predstavitvena datoteka, prenos (355,45 KB)
MD5: AB922B75137D67BEBAC24C303433D9CE
URLURL - Izvorni URL, za dostop obiščite https://iopscience.iop.org/article/10.1088/1361-6498/ae6081 Povezava se odpre v novem oknu

Izvleček
Occupational use of ionising radiation in medical procedures requires stringent protective measures to reduce exposure risks for staff. This study examined the integrity and reliability of personal protective equipment (PPE) used in a hospital setting, focusing on methods for detecting damage and its impact on radiation transmission. A total of 64 lead-based protective coats (one and two pieces) were examined for damage using four inspection techniques: visual- palpatory examination, LED light source, computed tomography (CT), and fluoroscopy/conventional x-ray. In the second part of the study, the transmittance of the primary beam and scattered radiation through damaged and undamaged areas was measured. Damage was found in 25% of protective coats, with 15.6%–18.8% classified as unsuitable according to recommended rejection models. Inspection using fluoroscopy and conventional x-ray was the most accurate method for detecting damage, followed by CT. Visual-palpatory, and light source assessments were less reliable, detecting 73.6% and 47.4% of all defects, respectively. Transmission of the primary beam in damaged areas increased from 0.5% to 316.6% compared to undamaged areas. Defective areas led to increased radiation exposure, potentially resulting in higher occupational doses and health concerns. The study emphasises the need for routine PPE inspections and recommends incorporating comprehensive assessments, including radiographic evaluations, into annual quality assurance and quality control protocols, as these have been shown to be most effective, particularly assessments using conventional x-ray units and fluoroscopy. These measures can significantly reduce the risk of unnecessary radiation exposure. However, time efficiency, required resources, and potential risks should also be considered and further evaluated.

Jezik:Angleški jezik
Ključne besede:radiation protection, PPE, lead shielding, protective coat, quality control
Vrsta gradiva:Članek v reviji
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:ZF - Zdravstvena fakulteta
Status publikacije:Objavljeno
Različica publikacije:Objavljena publikacija
Datum objave:01.01.2026
Leto izida:2026
Št. strani:8 str.
Številčenje:Vol. 46, no. 2, art. 021509
PID:20.500.12556/RUL-182217 Povezava se odpre v novem oknu
UDK:616-073
ISSN pri članku:1361-6498
DOI:10.1088/1361-6498/ae6081 Povezava se odpre v novem oknu
COBISS.SI-ID:276867843 Povezava se odpre v novem oknu
Datum objave v RUL:06.05.2026
Število ogledov:21
Število prenosov:7
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
Objavi na:Bookmark and Share

Gradivo je del revije

Naslov:Journal of radiological protection
Skrajšan naslov:J. radiol. prot.
Založnik:IOP Publishing
ISSN:1361-6498
COBISS.SI-ID:515075865 Povezava se odpre v novem oknu

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.
Začetek licenciranja:04.05.2026

Podobna dela

Podobna dela v RUL:
Podobna dela v drugih slovenskih zbirkah:

Nazaj